DocumentCode :
337201
Title :
An interface for nerve recording and stimulation with cuff electrodes
Author :
Sahin, Mesut ; Durand, Dominique M.
Author_Institution :
Appl. Neural Control Lab., Case Western Reserve Univ., Cleveland, OH, USA
Volume :
5
fYear :
1997
fDate :
1997
Firstpage :
2004
Abstract :
A nerve cuff electrode interface capable of both stimulating and recording from a nerve is described. The interface also rejects the EMG contamination in the recordings using reactive components without adding noise to the ENG signal. A transformer is added to the design for noise matching and the signal-to-noise ratio improvement is evaluated for a specific amplifier (AMP-O1)
Keywords :
biomedical electrodes; biomedical electronics; electromyography; neurophysiology; EMG contamination rejection; amplifier; cuff electrodes; nerve recording; nerve stimulation; noise matching; signal-to-noise ratio improvement; transformer; Biomedical electrodes; Contamination; Electromyography; Impedance; Noise cancellation; Noise figure; Noise generators; Signal design; Signal to noise ratio; Student members;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Engineering in Medicine and Biology Society, 1997. Proceedings of the 19th Annual International Conference of the IEEE
Conference_Location :
Chicago, IL
ISSN :
1094-687X
Print_ISBN :
0-7803-4262-3
Type :
conf
DOI :
10.1109/IEMBS.1997.758737
Filename :
758737
Link To Document :
بازگشت