• DocumentCode
    3373219
  • Title

    A synthetic method for improving magnetic resonance image contrast: Simulation and phantom study

  • Author

    Yun-Ni Ting ; Po-Chou Chen ; Cheng-Hsin Wang ; Herng-Hua Chang ; Woei-Chyn Chu

  • Author_Institution
    Inst. of Bio Med. Eng., Nat. Yang-Ming Univ., Taipei, Taiwan
  • fYear
    2013
  • fDate
    16-18 Dec. 2013
  • Firstpage
    96
  • Lastpage
    100
  • Abstract
    Magnetic resonance imaging has been widely applied in many clinical diagnoses. In order to increase the accuracy of diagnosis, it is crucial to improve the image quality. This paper proposed a method that modifies the bilateral filter with automatic parameter selection in local contrast correction (LCC) capable to adjust image contrast while maintaining high image quality in MRI. The proposed approach was validated by using both simulated images and MR images of an ACR phantom. The results confirmed that the proposed method outperforms several existing methods in both quantitative measurements and visual evaluations.
  • Keywords
    biomedical MRI; image enhancement; medical image processing; phantoms; ACR phantom; LCC; MRI; automatic parameter selection; bilateral filter; image quality; local contrast correction; magnetic resonance image contrast; phantom; Filtering; Image quality; Image restoration; Imaging phantoms; Magnetic resonance; PSNR; MRI; automatic biliteral filtering; local contrast correction;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Biomedical Engineering and Informatics (BMEI), 2013 6th International Conference on
  • Conference_Location
    Hangzhou
  • Print_ISBN
    978-1-4799-2760-9
  • Type

    conf

  • DOI
    10.1109/BMEI.2013.6746914
  • Filename
    6746914