Title :
Comparative analysis in terms of computational cost for different discrimination algorithms in implantable defibriilators
Author :
Cebrian, A. ; Rey, J. ; Millet, J. ; Rieta, J.J. ; Garcia-Civera, R.
Author_Institution :
Dept. of Electron. Eng., Polytech. Univ. of Valencia, Spain
Abstract :
Implantable defibrillators (ICDs) use very low computational cost criteria (rate, stability and onset) offering good sensitivity for arrhythmia detection. Although, the specificity of these combined criteria decreases in difficult arrhythmia discrimination as in case of discrimination between ventricular tachycardia (VT) and supraventricular tachycardia (SVT). Several morphological published algorithms enhance arrhythmia discrimination but most algorithms are developed in personal computers and cannot be used in ICDs because of computational cost requirements compared with limited ICD capabilities. A general method to determine the possibility of ICD implementation for a discrimination algorithm has been proposed.
Keywords :
defibrillators; diseases; medical computing; microcomputers; orthotics; ICD implementation; arrhythmia detection; arrhythmia discrimination; computational cost criteria; discrimination algorithm; good sensitivity; implantable defibrillators; morphological published algorithm; personal computers; supraventricular tachycardia; ventricular tachycardia; Algorithm design and analysis; Computational efficiency; Microcomputers; Morphology; Rhythm; Stability criteria; Testing; Time domain analysis; Wavelet analysis; Wavelet transforms;
Conference_Titel :
Computers in Cardiology, 2004
Print_ISBN :
0-7803-8927-1
DOI :
10.1109/CIC.2004.1442859