Title :
Measurement of left ventricular ejection fraction (EF) by densitometry from digital subtraction angiography
Author :
Machnig, Th ; Eicker, B. ; Barth, Kurt ; Lehmkuhl, H. ; Bachmann, K.
Author_Institution :
Div. of Cardiology, Erlangen Univ., Germany
Abstract :
The densitometric and geometric method for estimating the left ventricular ejection fraction (EF) are compared and the accuracy and reproducibility of densitometry in patients and casts are assessed. EF calculations for the densitometric method are in a better agreement with the simulated EF in the heart casts than the area-length method. The intra- and inter-observer variability is remarkably lower for the densitometry than for the area-length method. The data indicate that the densitometric calculation of EF is a more reliable method than the area-length method, because densitometry is independent of geometric assumptions
Keywords :
biomedical measurement; cardiology; densitometry; diagnostic radiography; area-length method; densitometry; digital subtraction angiography; geometric method; heart casts; left ventricular ejection fraction measurement; patients; Angiography; Biomedical engineering; Biomedical imaging; Cardiology; Computational modeling; Density measurement; Heart; Imaging phantoms; Shape; X-ray imaging;
Conference_Titel :
Computers in Cardiology 1990, Proceedings.
Conference_Location :
Chicago, IL
Print_ISBN :
0-8186-2225-3
DOI :
10.1109/CIC.1990.144287