• DocumentCode
    3381154
  • Title

    An Instrument for the Objective Measurement of Light Flicker

  • Author

    Gallo, Daniele ; Landi, Carmine ; Pasquino, Nicola

  • Author_Institution
    Dept. of Inf. Eng., Second Univ. of Naples, Aversa
  • Volume
    3
  • fYear
    2005
  • fDate
    16-19 May 2005
  • Firstpage
    1942
  • Lastpage
    1947
  • Abstract
    Measurement of light flicker with a low cost digital device, based on light intensity measurements, is proposed. Unlike current standard instrument based on IEC EN 61000-4-15, which determines flicker severity by filtering powering voltage measurements with a standard lamp model (230V/60W), the presented methodology measures directly the disturbances caused by light intensity variations including the lamp itself in the measurement process. In this way an objective measurement of light flicker is obtained. The metrological characterization of the device has been implemented by means of an automatic testing station which includes a standard IEC flickermeter (Norma D6000) as reference instrument. Thanks to its portability, the instrument can be used to measure light flicker independently of the lighting system, over a wide range of environments spanning from home/office to industrial locations
  • Keywords
    automatic testing; lamps; photometers; IEC flickermeter; automatic testing; light flicker measurement; light intensity measurements; low cost digital device; powering voltage measurement; standard lamp model; Automatic testing; Costs; Current measurement; Filtering; IEC standards; Instruments; Lamps; Measurement standards; Voltage fluctuations; Voltage measurement;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Instrumentation and Measurement Technology Conference, 2005. IMTC 2005. Proceedings of the IEEE
  • Conference_Location
    Ottawa, Ont.
  • Print_ISBN
    0-7803-8879-8
  • Type

    conf

  • DOI
    10.1109/IMTC.2005.1604510
  • Filename
    1604510