DocumentCode :
3384814
Title :
Study on the application of robust statistical techniques in interlaboratory comparisons for oil emission spectrometers
Author :
Pei, Wen ; Tian, Hong-Xiang ; Liu, Yu
Author_Institution :
Coll. of Naval Archit. & Power, Naval Univ. of Eng., Wuhan, China
Volume :
3
fYear :
2010
fDate :
9-10 Oct. 2010
Firstpage :
19
Lastpage :
23
Abstract :
In order to insure the reliability and the validity of the detection capability and detection data, 10 oil emission spectrometry laboratories which distribute in different areas tested 3 lubricating oil samples composed by different elements content, and interlaboratory comparisons for the detection capability of oil emission spectrometers were carried out. Based on robust statistical techniques, the detection data was evaluated and the evaluation results of robust Z-score were compared with the results of En. The results shows that the robust Z-score method had 86.67% hit rate, 7.33% failed rate, 6.00% omitted rate; while the En method had 31.33% hit rate, 66.67% failed rate, 2.00% omitted rate. the robust Z-score method is more suitable to be applied to interlaboratory comparisons for oil emission spectrometers than En method.
Keywords :
laboratory techniques; lubricating oils; spectrometers; statistical analysis; Z-score method; lubricating oil samples; oil emission spectrometers; robust statistical techniques; Copper; Iron; Lead; Nickel; Silicon; Tin; Zinc; En; interlaboratory comparisons; robust Z-score; spectrometer;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Future Information Technology and Management Engineering (FITME), 2010 International Conference on
Conference_Location :
Changzhou
Print_ISBN :
978-1-4244-9087-5
Type :
conf
DOI :
10.1109/FITME.2010.5654766
Filename :
5654766
Link To Document :
بازگشت