• DocumentCode
    3392046
  • Title

    A multiband concurrent sampling based RF front end for biotelemetry applications

  • Author

    Heragu, Aravind ; Balasubramanian, Viswanathan ; Enz, Christian

  • Author_Institution
    Integrated Syst. Lab., Ecole Polytech. Fed. de Lausanne (EPFL), Lausanne, Switzerland
  • fYear
    2010
  • fDate
    May 30 2010-June 2 2010
  • Firstpage
    2948
  • Lastpage
    2951
  • Abstract
    A sub-sampling based multiband RF front end suitable for medical telemetry applications is presented. Using a single inductor and a single sampling frequency it is shown how to downconvert concurrently Bluetooth LE and MICS band signals to baseband with relative attenuation of 42 dB between the two downconverted bands. The proposed front end is analyzed in detail and the results are validated using Spectre RF simulations for a 0.18μm CMOS process.
  • Keywords
    Bluetooth; CMOS integrated circuits; medical signal processing; signal sampling; telemetry; Bluetooth LE; CMOS process; MICS band signals; Spectre RF simulations; biotelemetry applications; inductor; medical telemetry applications; multiband concurrent sampling; size 0.18 mum; sub-sampling based multiband RF front end; Analytical models; Attenuation; Baseband; Biomedical telemetry; Bluetooth; Inductors; Microwave integrated circuits; Radio frequency; Sampling methods; Signal sampling;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Circuits and Systems (ISCAS), Proceedings of 2010 IEEE International Symposium on
  • Conference_Location
    Paris
  • Print_ISBN
    978-1-4244-5308-5
  • Electronic_ISBN
    978-1-4244-5309-2
  • Type

    conf

  • DOI
    10.1109/ISCAS.2010.5538036
  • Filename
    5538036