• DocumentCode
    3398663
  • Title

    Electric fields in the heart in internal defibrillation

  • Author

    Shankar, B. ; Kothiyal, K.P. ; Fogelson, L. ; Thakor, Nitish V.

  • Author_Institution
    Dept. of Biomed. Eng., Johns Hopkins Sch. of Med., Baltimore, MD, USA
  • fYear
    1988
  • fDate
    4-7 Nov. 1988
  • Firstpage
    183
  • Abstract
    The automatic implantable defibrillator delivers a shock directly to the heart. Optimal strategies for delivering these shocks were determined by studying a three-dimensional computer model of the electric fields produced by internal defibrillation electrodes. A finite-element technique was used to determine energy and current distribution in the heart for electrode configurations typically in use clinically. The model provides useful data on regional current and energy distributions.<>
  • Keywords
    bioelectric phenomena; cardiology; defibrillators; automatic implantable defibrillator; electric fields; finite-element technique; heart; three-dimensional computer model;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 1988. Proceedings of the Annual International Conference of the IEEE
  • Conference_Location
    New Orleans, LA, USA
  • Print_ISBN
    0-7803-0785-2
  • Type

    conf

  • DOI
    10.1109/IEMBS.1988.94468
  • Filename
    94468