Title :
Response characteristics of the TlBr radiation detector as a function of the detector temperature
Author :
Shoji, T. ; Hitomi, K. ; Suehiro, T. ; Hiratate, Y.
Author_Institution :
Dept. of Electron., Tohoku Inst. of Technol., Sendai, Japan
Abstract :
The radiation detectors were fabricated from the TlBr crystals grown by TMZ method and the FWHM and transit time of electrons and holes were measured as a function of the detector temperature. The TlBr radiation detector shows the best response characteristics at about 320 K for holes. At the temperature of 330 K or above, the transit time and FWHM for the 241Am α-particle peak become worse. On the other hand, in the case of electron, the spectrum becomes worse with increasing the temperature form 150 K to 380 K. In the case, we can deduce an activation energy of about 0.20 eV for holes and about 0.10 for electron from the temperature dependence of the carrier transit time
Keywords :
alpha-particle detection; electron detection; semiconductor counters; thallium compounds; α-particle peak; 100 keV to 1 MeV; 220 to 390 K; TlBr; activation energy; detector temperature; electron detection; electron transit time; hole transit time; response characteristics; semiconductor radiation detector; temperature dependence; Absorption; Charge carrier processes; Crystalline materials; Crystals; Electrodes; Furnaces; Pulse measurements; Radiation detectors; Temperature dependence; Time measurement;