Retinal pulse oximetry: towards a method for measuring cerebral oxygen saturation
Author
Scott, Valerie A. ; Tarassenko, Lionel ; Glynn, Christopher J. ; Hill, Adrian R.
Author_Institution
Dept. of Eng. Sci., Oxford Univ., UK
Volume
2
fYear
1995
fDate
20-23 Sep 1995
Firstpage
1555
Abstract
An instrument to make reflectance pulse oximetry measurements on the blood in the retinal fundus has been developed. Conventional pulse oximeters measure the oxygen saturation of arterial blood in the peripheral circulation, whereas using the retina provides the opportunity to monitor cerebral oxygenation. Cardiac-synchronous pulsations have been observed in the red and IR signals during preliminary in vivo trials, however it is unclear whether these pulsations are entirely due to the pulsating volume of arterial blood. Analysis of a series of retinal images has revealed that pulsations can be detected from the veins and choroidal background as well as from the arteries
Keywords
biomedical equipment; biomedical measurement; blood; brain; chemical variables measurement; eye; oxygen; IR signals; O2; arterial blood pulsating volume; arteries; cardiac-synchronous pulsations; cerebral oxygen saturation measurement; choroidal background; medical instrument; peripheral circulation; preliminary in vivo trials; reflectance pulse oximetry measurements; retinal images series analysis; retinal pulse oximetry; veins; Arterial blood circulation; Arteries; Image analysis; In vivo; Instruments; Monitoring; Pulse measurements; Reflectivity; Retina; Veins;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society, 1995., IEEE 17th Annual Conference
Conference_Location
Montreal, Que.
Print_ISBN
0-7803-2475-7
Type
conf
DOI
10.1109/IEMBS.1995.579824
Filename
579824
Link To Document