Title :
Development of an instrumented walker for measurement of unilateral hand loads
Author :
Bachschmidt, R.A. ; Harris, Gerald F. ; Simoneau, Guy G.
Author_Institution :
Dept. of Biomed. Eng., Marquette Univ., Milwaukee, WI, USA
Abstract :
The purpose of this study was to design and develop an instrumentation configuration suitable for determining upper extremity forces during walker-assisted gait. To accomplish these goals a finite element model of a conventional aluminum tube walker was constructed and the structural cross-talk in frame members was fully described for orthogonal loading conditions. Results from the study were used to locate strain gage sensors on the structure such that independent measurements of unilateral hand loads could be determined. The instrumentation configuration is presented along with calibration data illustrating system output. Complete description of upper extremity kinetics during walker-assisted gait may provide insight into rehabilitative treatment strategies. Characterization of the demands placed upon the upper extremities and assessment of patient stability may influence walker design parameters (height, width, weight) and recommended procedures for use (foot placement, hand location, motion sequence, and feedback loading control)
Keywords :
biomechanics; calibration; finite element analysis; handicapped aids; patient treatment; strain gauges; aluminum tube walker; calibration data; feedback loading control; finite element model; foot placement; frame members; hand location; height; instrumentation configuration; instrumented walker; motion sequence; orthogonal loading conditions; patient stability; rehabilitative treatment strategies; strain gage sensors; structural cross-talk; system output; unilateral hand loads; upper extremities; upper extremity forces; upper extremity kinetics; walker design parameters; walker-assisted gait; weight; width; Aluminum; Calibration; Capacitive sensors; Extremities; Finite element methods; Instruments; Kinetic theory; Sensor phenomena and characterization; Stability; Strain measurement;
Conference_Titel :
Biomedical Engineering Conference, 1996., Proceedings of the 1996 Fifteenth Southern
Conference_Location :
Dayton, OH
Print_ISBN :
0-7803-3131-1
DOI :
10.1109/SBEC.1996.493111