DocumentCode :
3404046
Title :
Uncertainty evaluation on the freezing point of silver for thermocouple calibration
Author :
Numajiri, Haruhiko ; Ogura, Hideki ; Izuchi, Masaya ; Arai, Masaru
Author_Institution :
Nat. Inst. of Adv. Ind. Sci. & Technol., Nat. Metrol. Inst. of Japan, Ibaraki, Japan
Volume :
2
fYear :
2002
fDate :
5-7 Aug. 2002
Firstpage :
1037
Abstract :
The freezing point of silver is realized using an open type cell. The uncertainty of the thermocouple calibration of this temperature was evaluated using Pt/Pd thermocouples. The sources of uncertainty estimated were stability and repeatability of the plateau, temperature distribution in the cell. impurity effect of the metal, surroundings pressure correction and hydrostatic pressure correction. The combined standard uncertainty is 7.8 mK.
Keywords :
calibration; thermocouples; uncertainty handling; calibration; combined standard uncertainty; freezing point; silver; thermocouples; uncertainty evaluation; Ambient intelligence; Atmospheric measurements; Calibration; Furnaces; Insulation; Resistance heating; Silver; Stability; Temperature; Uncertainty;