Title :
Wireless and batteryless blood pressure sensor
Author :
Sosa, J. ; Garcia-Alonso, S. ; Monzon-Verona, J.M. ; Montiel-Nelson, J.A. ; Beriain, A. ; Fernandez, Eduardo ; Solar, H. ; Garcia-Alonso, A. ; Gutierrez, I.
Author_Institution :
Dept. of Electron. Eng. & Automatics, Univ. of Las Palmas de Gran Canaria, Las Palmas de Gran Canaria, Spain
Abstract :
A low-power invasive blood pressure transducer has been designed. It is based on a structure of four high resistive strain gauges in a full Wheatstone bridge configuration. For obtaining the required sensitivity, the gauge was allocated performing a careful mechanical simulation of the membrane behaviour. The electronic conditioning circuit and the ADC were optimized to compensate the noise introduced by the high resistive gauges. A batteryless RFID tag designed to incorporate and to supply the sensor was fabricated and measured. It power supplies the sensor at a distance longer than 50 cm from a standard reader, while the reported ranges -to the best of our knowledge- are shorter than 10 cm. This wireless and batteryless sensor is especially suitable for surgery rooms and intensive care units.
Keywords :
analogue-digital conversion; blood pressure measurement; pressure sensors; radiofrequency identification; surgery; ADC; Wheatstone bridge; batteryless RFID tag; batteryless blood pressure sensor; electronic conditioning circuit; high resistive strain gauges; intensive care units; low-power invasive blood pressure transducer; mechanical simulation; membrane behaviour; standard reader; surgery rooms; wireless blood pressure sensor; Art; Lead;
Conference_Titel :
Circuits and Systems (MWSCAS), 2011 IEEE 54th International Midwest Symposium on
Conference_Location :
Seoul
Print_ISBN :
978-1-61284-856-3
Electronic_ISBN :
1548-3746
DOI :
10.1109/MWSCAS.2011.6026397