DocumentCode :
3407583
Title :
Prevention of left atrial wall suction in a continuous-flow total artificial heart by controlling right pump output
Author :
Olegario, P.S. ; Yoshizawa, M. ; Tanaka, A. ; Abe, K. ; Yambe, T. ; Nitta, S. ; Takeda, H.
Author_Institution :
Tohoku Univ., Sendai, Japan
Volume :
5
fYear :
2002
fDate :
5-7 Aug. 2002
Firstpage :
2855
Abstract :
One of the problems in continuous-flow total artificial heart control is atrial wall suction. In this study we developed a control algorithm which uses the right pump to prevent left atrial wall suction from occurring. Results of in vitro and in vivo tests indicate that prevention of left atrial wall suction is possible using this method and that it can be performed simultaneously with physiological control of the artificial heart.
Keywords :
artificial organs; cardiology; haemodynamics; medical control systems; artificial hearts control; atrial suction; automatic control; continuous-flow; in vitro tests; in vivo tests; physiological control; right pump; total artificial heart; Algorithm design and analysis; Animal structures; Artificial heart; Automatic control; Blood; Control systems; Energy efficiency; Humans; Pressure control; Radio control;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
SICE 2002. Proceedings of the 41st SICE Annual Conference
Print_ISBN :
0-7803-7631-5
Type :
conf
DOI :
10.1109/SICE.2002.1195552
Filename :
1195552
Link To Document :
بازگشت