Title :
Comparison measurement of water triple point cells for uncertainty evaluation
Author :
Kishimoto, I. ; Sato, K. ; Koizumi, K. ; Arai, M.
Author_Institution :
Nat. Metrol. Inst. of Japan, Nat. Inst. of Adv. Ind. Sci. & Technol., Ibaraki, Japan
Abstract :
Comparison measurement was performed on some triple point of water (TPW) cells, including national standard TPW cells. The purpose is to reassess the uncertainty source of realized temperatures of the TPW cells. The results showed the largest difference among them is represented as 0.036 ± 0.035 mK with the coverage factor k = 2.
Keywords :
critical points; measurement uncertainty; resistance thermometers; temperature measurement; temperature scales; comparison measurement; coverage factor; national standard TPW cells; standard platinum resistance thermometer; temperature standard; triple point of water cells; uncertainty evaluation; water triple point cells; Calibration; Electrical resistance measurement; Ice; Impurities; Manufacturing; Measurement uncertainty; Performance evaluation; Scattering; Temperature measurement; Thermal resistance;
Conference_Titel :
SICE 2002. Proceedings of the 41st SICE Annual Conference
Print_ISBN :
0-7803-7631-5
DOI :
10.1109/SICE.2002.1195565