• DocumentCode
    3407860
  • Title

    Non-uniqueness of the International Temperature Scale of 1990 in the range 14 K to 433 K

  • Author

    Sakurai, H.

  • Author_Institution
    Nat. Inst. of Adv. Ind. Sci. & Technol., Japan
  • Volume
    5
  • fYear
    2002
  • fDate
    5-7 Aug. 2002
  • Firstpage
    2921
  • Abstract
    The non-uniqueness of the International Temperature Scale of 1990 is estimated using 6 standard capsule type platinum resistance thermometers in the range from 14 K to 302 K and 7 standard long-stem type platinum thermometers in the range 0°C to 156°C. The non-uniqueness is found to be less than 0.3 mK in almost all ranges, except a few subranges that cover from 14 K to 54 K. The inconsistency between subranges is larger than the expected value. Both must be included in uncertainty sources in resistance thermometry.
  • Keywords
    calibration; critical points; measurement uncertainty; resistance thermometers; temperature scales; 0 to 156 C; 14 to 433 K; ITS-90; International Temperature Scale of 1990; capsule type platinum resistance thermometers; inconsistency; long-stem type platinum thermometers; nonuniqueness; resistance thermometry; triple point of water; uncertainty sources; Electrical resistance measurement; Equations; Interpolation; Measurement standards; Platinum; Temperature control; Temperature distribution; Temperature measurement; Thermal resistance; Wire;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    SICE 2002. Proceedings of the 41st SICE Annual Conference
  • Print_ISBN
    0-7803-7631-5
  • Type

    conf

  • DOI
    10.1109/SICE.2002.1195567
  • Filename
    1195567