• DocumentCode
    3407870
  • Title

    Study of preparation of ice mantles of triple point of water cells

  • Author

    Kodaira, K.

  • Author_Institution
    Japan Electr. Meters Inspection Corp., Tokyo, Japan
  • Volume
    5
  • fYear
    2002
  • fDate
    5-7 Aug. 2002
  • Firstpage
    2927
  • Abstract
    Ice mantles of triple point of water cells were prepared by a new technique using a remodeled refrigerator with a circulation pump. The work was completed within approximately 10 minutes. No cracks were formed in ice mantles during the preparation.
  • Keywords
    calibration; critical points; ice; resistance thermometers; specimen preparation; temperature scales; ITS-90; Kelvin Thermodynamic Temperature Scale; calibration; circulation pump; ice mantle preparation; remodeled refrigerator; standard platinum resistance thermometers; triple point of water cells; Cooling; Ethanol; Ice; Nitrogen; Ocean temperature; Refrigeration; Solids; Thermodynamics; Valves; Watthour meters;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    SICE 2002. Proceedings of the 41st SICE Annual Conference
  • Print_ISBN
    0-7803-7631-5
  • Type

    conf

  • DOI
    10.1109/SICE.2002.1195568
  • Filename
    1195568