Title :
New approach to calibration in PD measurements
Author :
Carminati, E. ; Lazzaroni, M.
Author_Institution :
Dipt. di Elettrotecnica, Politecnico di Milano, Italy
Abstract :
Two different PD measurement instruments could detect different PD values on the same test object even if the measurement conditions are identical. This result is due to the quasi integration error presents in classical PD measurement systems. In this paper the quasi integration error is investigated and two parameters are proposed to qualify the PD detection instrumentation. Moreover a new approach to instrument calibration is depicted
Keywords :
calibration; insulation testing; integrating circuits; measurement errors; partial discharge measurement; DAC; HV test object insulation; PD measurements; analog integrator; calibration; current pulses; ideal integration; measurement instruments; quasi integration error; square wave generator; traceability; Calibration; Charge measurement; Circuit testing; Current measurement; Instruments; Integrated circuit measurements; Partial discharges; Pulse generation; Pulse measurements; Voltage;