DocumentCode
3415016
Title
Computer aided analysis of the radio-frequency birdcage resonator
Author
Garcia, B.J. ; Pascone, R.J. ; Fitzgerald, T.M.
Author_Institution
Dept. of Electr. Eng., Manhattan Coll., Riverdale, NY, USA
fYear
1988
fDate
4-7 Nov. 1988
Firstpage
392
Abstract
Computer analysis has been used to investigate electrical and physical properties of the birdcage coil used in MRI (magnetic resonance imaging). These studies are based on electrical circuit analysis using SPICE software and on Biot-Savart calculations of the radiated RF field distribution. Significant results include verification of resonance mode peaks according to the Brillouin formula and determination of a nonlinear dependence of RF field homogeneity on number of column elements in the coil. These studies are aimed at providing an improved theoretical understanding of the relation between performance and the electrical and physical parameters of the MRI coil device.<>
Keywords
biomedical NMR; biomedical equipment; coils; computer aided analysis; medical computing; resonators; Biot-Savart calculations; Brillouin formula; RF birdcage resonator; RF field homogeneity; SPICE software; coil column elements; computer aided analysis; electrical circuit analysis; electrical properties; magnetic resonance imaging; physical properties;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society, 1988. Proceedings of the Annual International Conference of the IEEE
Conference_Location
New Orleans, LA, USA
Print_ISBN
0-7803-0785-2
Type
conf
DOI
10.1109/IEMBS.1988.94573
Filename
94573
Link To Document