DocumentCode
3424311
Title
Manipulation of whole blood using traveling wave dielectrophoresis
Author
Lo, U. ; Wo, A.M. ; Lei, U.
Author_Institution
Inst. of Appl. Mech., Nat. Taiwan Univ., Taipei, Taiwan
fYear
2005
fDate
30 Jan.-3 Feb. 2005
Firstpage
686
Lastpage
689
Abstract
This paper presents new results on manipulation of undiluted whole blood using traveling wave dielectrophoresis (twDEP). Although twDEP is not new, data on whole blood manipulation is rare. The twDEP in this work consists of four 2D electrodes of 20μm width operating at 10MHz sinusoidal wave with 90 degree phase apart. Results show that whole blood bulk flow can be manipulated at-will in either direction. Moreover, simultaneous bi-direction motion is also observed at different height of the 25 μm flow channel. This flow feature might be utilized for separation purpose.
Keywords
biological techniques; biomedical measurement; blood; channel flow; electrophoresis; microfluidics; patient diagnosis; 10 MHz; 2D electrodes; bidirection motion; blood bulk flow; blood manipulation; flow channel; sinusoidal wave; traveling wave dielectrophoresis; whole blood; Bidirectional control; Blood; Cells (biology); Dielectrics; Dielectrophoresis; Electrodes; Frequency; Nonuniform electric fields; Permittivity; Proteins;
fLanguage
English
Publisher
ieee
Conference_Titel
Micro Electro Mechanical Systems, 2005. MEMS 2005. 18th IEEE International Conference on
ISSN
1084-6999
Print_ISBN
0-7803-8732-5
Type
conf
DOI
10.1109/MEMSYS.2005.1454022
Filename
1454022
Link To Document