Title :
Development of an SU-8 Fabry-Perot blood pressure sensor
Author :
Melamud, R. ; Davenport, A.A. ; Hill, G.C. ; Chan, I.H. ; Declercq, F. ; Hartwell, P.G. ; Pruitt, B.L.
Author_Institution :
Dept. of Mech. Eng., Stanford Univ., CA, USA
fDate :
30 Jan.-3 Feb. 2005
Abstract :
This paper presents the fabrication method and testing of an interferometric pressure sensor designed for intravascular blood pressure measurements. A cap containing a pressure-sensing diaphragm was mounted onto the end of a fiber optic cable. The Fabry-Perot interferometer, formed between the reflective diaphragm and the fiber´s end, measured the diaphragm deflection. Microfabricated from the biocompatible polymer, SU-8, the device is fast, simple, and inexpensive to manufacture. Its small dimensions (<300μm outer diameter) reduce the risk of infection and thrombosis and allow for its insertion into small vessels. The sensor showed a linear response to pressure from 0 to 125 mmHg with approximately 1-2 mmHg resolution. The use of an optical displacement transducer allowed a series of careful measurements of drift and hysteresis of SU-8 materials in different environments. This data may be valuable to other researchers working with SU-8.
Keywords :
Fabry-Perot interferometers; biosensors; blood pressure measurement; microsensors; pressure sensors; 0 to 125 mmHg; SU-8 Fabry-Perot blood pressure sensor; biocompatible polymer; diaphragm deflection; fiber optic cable; hysteresis; interferometric pressure sensor; intravascular blood pressure measurement; optical displacement transducer; pressure-sensing diaphragm; thrombosis; Biosensors; Blood pressure; Fabry-Perot interferometers; Optical device fabrication; Optical fiber cables; Optical fiber devices; Optical fibers; Optical interferometry; Pressure measurement; Testing;
Conference_Titel :
Micro Electro Mechanical Systems, 2005. MEMS 2005. 18th IEEE International Conference on
Print_ISBN :
0-7803-8732-5
DOI :
10.1109/MEMSYS.2005.1454053