Abstract :
Storing digital medical images is standardized by the DICOM report. Lossy pulse-echo ultrasonic image compression by a JPEG baseline system is permitted by it. The purpose of this paper is to reduce the data volume and to achieve a low bit rate in the digital representation of pulse-echo ultrasonic images without perceived loss of image quality. In image compression with a JPEG baseline system, it is possible to control the compression ratio and image quality by controlling quantization values. This paper attempts standardizing JPEG quantization table for medical ultrasonic echo images by a statistical method. Results reveal that the proposed method achieves a lower bit rate compared with the JPEG standard under the same image quality
Keywords :
biomedical ultrasonics; data compression; echo; medical image processing; quantisation (signal); statistical analysis; JPEG baseline system; JPEG quantization table; bit rate; compression ratio; data volume; digital medical images; digital representation; image quality; lossy pulse-echo ultrasonic image compression; medical ultrasonic echo images; statistical method; Biomedical imaging; Bit rate; Control systems; DICOM; Image coding; Image quality; Medical control systems; Quantization; Standardization; Transform coding;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Electronics, Circuits and Systems, 1999. Proceedings of ICECS '99. The 6th IEEE International Conference on
Conference_Location :
Pafos
Print_ISBN :
0-7803-5682-9
Type :
conf
DOI :
10.1109/ICECS.1999.813201
Filename :
813201
Link To Document :
بازگشت