DocumentCode
3430128
Title
An implantable cuff electrode for collision block of pudendal nerve motor activity
Author
Mortimer, J. Thomas ; Sweeney, James D. ; Bodner, Donald R. ; Ferguson, A. Stewart
Author_Institution
Case Western Reserve Univ., Cleveland, OH, USA
fYear
1988
fDate
4-7 Nov. 1988
Firstpage
1523
Abstract
An implantable cuff electrode technique for collision block of pudendal nerve motor activity has been developed and tested through acute and chronic animal studies. The necessary quasitrapezoidal stimulus parameters for acute and chronic collision block generation have been characterized and are discussed. An asymmetric two-electrode cuff design effectively produces the trains of antidromically propagated action potentials needed to implement a collision block of motor signals to the external periurethral sphincter.<>
Keywords
biomedical equipment; electrodes; neurophysiology; patient treatment; animal studies; antidromically propagated action potentials; collision block; external periurethral sphincter; implantable cuff electrode; medical implant; pudendal nerve motor activity; quasitrapezoidal stimulus parameters; therapeutic technique;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society, 1988. Proceedings of the Annual International Conference of the IEEE
Conference_Location
New Orleans, LA, USA
Print_ISBN
0-7803-0785-2
Type
conf
DOI
10.1109/IEMBS.1988.94659
Filename
94659
Link To Document