• DocumentCode
    3439997
  • Title

    4-subject 4-channel optical telemetry system for use in electrocardiograms

  • Author

    Park, Jongdae ; Son, Jinwoo ; Seo, Heedon ; Ishida, M.

  • Author_Institution
    Electron. & Telecommun. Res. Inst., Taejon, South Korea
  • Volume
    3
  • fYear
    1998
  • fDate
    1998
  • Firstpage
    251
  • Abstract
    The implementation of a 4-subject, 4-channel optical telemetry system for the measurement of electrocardiograms (ECGs) is described in this paper, a system which receives and transmits physiological signals to the external system using LED and infrared light of photodiode. This system decreases the dependency of power supply voltage to the CMOS IC chip and also uses a new enforced synchronization technique using infrared bi-directional communication. The telemeter IC with the size of 5.1×5.1 mm2 has the following functions: receiving of command signal, initialization of internal state of all functional blocks, decoding of subject selection signal, time division multiplexing of 4-channel modulated physiological signals, transmission of modulated signals to external system, and auto power down control. ECG is tested to confirm telemetry system
  • Keywords
    biomedical telemetry; electrocardiography; synchronisation; time division multiplexing; electrocardiograms; infrared bi-directional communication; optical telemetry system; power supply voltage; subject selection signal; synchronization technique; time division multiplexing; Bidirectional control; CMOS integrated circuits; Electrocardiography; Light emitting diodes; Optical fiber communication; Photodiodes; Power supplies; Semiconductor device measurement; Telemetry; Voltage;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Electronics, Circuits and Systems, 1998 IEEE International Conference on
  • Conference_Location
    Lisboa
  • Print_ISBN
    0-7803-5008-1
  • Type

    conf

  • DOI
    10.1109/ICECS.1998.813977
  • Filename
    813977