DocumentCode :
3443821
Title :
Investigation of cryogenic treatment of UHMWPE
Author :
Trieu, H.H. ; Morris, L.H. ; Kaufman, M.E. ; Hood, R. ; Jenkins, L.S.
Author_Institution :
Wright Med. Technol. Inc., Arlington, TN, USA
fYear :
1997
fDate :
4-6 Apr 1997
Firstpage :
90
Lastpage :
91
Abstract :
UHMWPE has been used as a bearing material in total joint arthroplasty for many years. Wear of UHMWPE can adversely affect the performance and longevity of orthopaedic implants. Various efforts have been focused on the improvement of UHMWPE properties, including wear resistance. In this study, a cryogenic treatment was investigated for potential improvements of UHMWPE. The cryogenic treatment applied in this study did not appear to induce any significant or detectable changes in UHMWPE. Wear testing of treated UHMWPE was not pursued based on the findings of this evaluation
Keywords :
bone; cryogenics; prosthetics; wear resistance; bearing material; cryogenic treatment; implant performance; orthopaedic implants longevity; prosthetic joints; total joint arthroplasty; wear resistance improvement; Cryogenics; Crystallization; Educational institutions; Implants; Joining materials; Optical microscopy; Polyethylene; Polymers; Temperature; Testing;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Biomedical Engineering Conference, 1997., Proceedings of the 1997 Sixteenth Southern
Conference_Location :
Biloxi, MS
ISSN :
1086-4105
Print_ISBN :
0-7803-3869-3
Type :
conf
DOI :
10.1109/SBEC.1997.583222
Filename :
583222
Link To Document :
بازگشت