DocumentCode
3446291
Title
Intercomparison of Precision Current Sources in the Range From 10 fA to 10 pA Between PTB and VNIIM
Author
Willenberg, G.D. ; Tauscher, H.N. ; Pavlov, O.M. ; Korotkova, I.
Author_Institution
Phys.-Technische Bundesanslalt, Braunschweig
fYear
2004
fDate
38139
Firstpage
256
Lastpage
256
Abstract
An intercomparison of precision current sources was carried out for currents between 10 fA to 10 pA. The current sources of both PTB and VNIIM were based on the principle of charging a capacitor using a linear voltage ramp (Rozhdestvenskaya et al., 1979) and (Willenberg et al., 2003). The relative deviation for the values of the sources were found to be between 2.8middot10-4 and 3.1middot10-3, depending on the current range
Keywords
capacitors; electric current measurement; measurement standards; 10 fA; 10 pA; PTB; VNIIM; capacitor charging; linear voltage ramp; precision current sources intercomparison; relative deviation; Calibration; Capacitors; Current measurement; Gain measurement; Instruments; Measurement standards; Metrology; Performance evaluation; Standards development; Voltage;
fLanguage
English