• DocumentCode
    3446291
  • Title

    Intercomparison of Precision Current Sources in the Range From 10 fA to 10 pA Between PTB and VNIIM

  • Author

    Willenberg, G.D. ; Tauscher, H.N. ; Pavlov, O.M. ; Korotkova, I.

  • Author_Institution
    Phys.-Technische Bundesanslalt, Braunschweig
  • fYear
    2004
  • fDate
    38139
  • Firstpage
    256
  • Lastpage
    256
  • Abstract
    An intercomparison of precision current sources was carried out for currents between 10 fA to 10 pA. The current sources of both PTB and VNIIM were based on the principle of charging a capacitor using a linear voltage ramp (Rozhdestvenskaya et al., 1979) and (Willenberg et al., 2003). The relative deviation for the values of the sources were found to be between 2.8middot10-4 and 3.1middot10-3, depending on the current range
  • Keywords
    capacitors; electric current measurement; measurement standards; 10 fA; 10 pA; PTB; VNIIM; capacitor charging; linear voltage ramp; precision current sources intercomparison; relative deviation; Calibration; Capacitors; Current measurement; Gain measurement; Instruments; Measurement standards; Metrology; Performance evaluation; Standards development; Voltage;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements Digest, 2004 Conference on
  • Conference_Location
    London
  • Print_ISBN
    0-7803-8494-6
  • Electronic_ISBN
    0-7803-8494-6
  • Type

    conf

  • DOI
    10.1109/CPEM.2004.305559
  • Filename
    4097214