DocumentCode :
3447040
Title :
Comparison of Two Impulse Calibrators with a High Resolution Digitizer
Author :
Li, Y. ; Hallstrom, J. ; Lucas, W.
Author_Institution :
National Meas. Lab., Lindfield, NSW
fYear :
2004
fDate :
38139
Firstpage :
326
Lastpage :
327
Abstract :
The lightning impulse voltage calibrators, developed at NML and MIKES-HUT, and a recently developed commercial digitiser were compared. With this three-way comparison, not only the accuracies of the devices were compared, but also the linearity of both peak voltage and time parameters of the devices was investigated. The results demonstrated that the linearity of the calibrators is superior to that of the digitiser
Keywords :
analogue-digital conversion; calibration; pulse generators; voltage measurement; high resolution digitizer; lightning impulse voltage calibrators; Attenuators; Australia; Circuits; DC generators; Impedance; Laboratories; Linearity; Measurement standards; Time measurement; Voltage measurement;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Precision Electromagnetic Measurements Digest, 2004 Conference on
Conference_Location :
London
Print_ISBN :
0-7803-8494-6
Electronic_ISBN :
0-7803-8494-6
Type :
conf
DOI :
10.1109/CPEM.2004.305597
Filename :
4097252
Link To Document :
بازگشت