DocumentCode :
3447199
Title :
An Automated System for the Precision Calibration of High Value Resistance Standards using a Novel Scanner
Author :
Rigby, R.G.
Author_Institution :
National Meas. Lab., CSIRO, Lindfield, NSW
fYear :
2004
fDate :
38139
Firstpage :
342
Lastpage :
343
Abstract :
A fully automated system for the precision calibration of high value resistance standards of 100 MOmega, and greater has been built and is in use at NML Australia. The precision obtained is due to the use of a novel scanner for the selection of reference and unknown resistance standards, the temperature control of these, and the characterisation of the comparator used, a commercially available teraohmmeter
Keywords :
calibration; computerised instrumentation; measurement standards; ohmmeters; 100 Mohm; commercially available teraohmmeter; comparator characterisation; fully automated system; high value resistance standards; precision calibration; reference selection; temperature control; Australia; Calibration; Circuits; Coaxial components; Digital relays; Electrical resistance measurement; Measurement standards; Petroleum; Resistors; Voltage;