DocumentCode :
3447529
Title :
Evaluation of the Frequency Dependence of the Resistance and Capacitance Standards in the BIPM Quadrature Bridge
Author :
Delahaye, F. ; Goebel, R.
Author_Institution :
Bur. Int. des Poids et Mesures, Sevres
fYear :
2004
fDate :
38139
Firstpage :
372
Lastpage :
373
Abstract :
The BIPM uses a quadrature bridge to link capacitance standards to resistance standards. The frequency dependence of the resistances used in the quadrature bridge (a pair of 51.6 kOmega resistors) is determined by comparison with a calculable coaxial resistor of 1290.6 Omega. The quadrature bridge can be operated at three different frequencies (1027 Hz, 1541 Hz and 3082 Hz) using three different pairs of capacitors of values 3000 pF, 2000 pF and 1000 pF, respectively. The frequency dependence of each capacitance in this range of frequencies is deduced from these measurements
Keywords :
capacitance measurement; electric resistance measurement; measurement standards; resistors; 1027 Hz; 12.90 ohm; 1541 Hz; 3082 Hz; 51.6 kohm; BIPM; capacitance standards; coaxial resistor; frequency dependence; quadrature bridge; resistance standards; Bridge circuits; Calibration; Capacitance measurement; Capacitors; Coaxial components; Electrical resistance measurement; Frequency dependence; Frequency measurement; Measurement standards; Resistors;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Precision Electromagnetic Measurements Digest, 2004 Conference on
Conference_Location :
London
Print_ISBN :
0-7803-8494-6
Electronic_ISBN :
0-7803-8494-6
Type :
conf
DOI :
10.1109/CPEM.2004.305620
Filename :
4097275
Link To Document :
بازگشت