    Impedance Bridge Ratio Transformer Calibration by Reference-Step Conductance Build-Up

  • Author

    Corney, A.C.

  • Author_Institution
    Meas. Stand. Lab., New Zealand Ind. Res. Ltd., Lower Hutt
  • fYear
    2004
  • fDate
    38139
  • Firstpage
    380
  • Lastpage
    381
  • Abstract
    A reference-step conductance build-up method suitable for automation has been developed for calibrating the ratio dial transformer of a wide-range impedance bridge. The uncertainties over the range 40 Hz to 20 kHz are estimated not to exceed 0.2 muV/V referred to the divider input voltage. At 100 kHz, the uncertainties do not exceed 2muA7V
  • Keywords
    calibration; electric admittance measurement; estimation theory; measurement uncertainty; transformers; 40 to 20000 Hz; impedance bridge; ratio dial transformer; ratio transformer calibration; reference step conductance build up; uncertainties; Admittance; Bridge circuits; Calibration; Frequency conversion; Impedance; Measurement standards; Power transformer insulation; Resistors; Switches; Voltage;