DocumentCode
3448553
Title
High Temperature Coaxial Line Measurement Cell using Room Temperature Calibration
Author
Le Gallou, J.-H. ; Ledieu, M. ; Acher, O.
Author_Institution
CEA, Monts
fYear
2004
fDate
38139
Firstpage
473
Lastpage
474
Abstract
This paper describes another use of the coaxial line technique for the measurement of the permittivity and permeability of magnetic and dielectric materials at elevated temperatures. The proposed modified coaxial measurement cell allows making fast and accurate broadband measurements taking advantage of a single room-temperature calibration. Measurements are performed from 50 MHz up to 20 GHz. The new coaxial cell is experimentally validated. Results for glass and ferrite samples, in the 20-300 degC temperature range, are presented
Keywords
coaxial cables; dielectric materials; high-temperature electronics; magnetic materials; magnetic permeability measurement; microwave measurement; permittivity measurement; 20 to 300 C; 50 to 20000 MHz; coaxial line measurement cell; dielectric materials; magnetic materials; permeability measurement; permittivity measurement; room-temperature calibration; Calibration; Coaxial components; Dielectric materials; Dielectric measurements; Ferrites; Glass; Performance evaluation; Permeability measurement; Permittivity measurement; Temperature measurement;
fLanguage
English
Publisher
ieee
Conference_Titel
Precision Electromagnetic Measurements Digest, 2004 Conference on
Conference_Location
London
Print_ISBN
0-7803-8494-6
Electronic_ISBN
0-7803-8494-6
Type
conf
DOI
10.1109/CPEM.2004.305316
Filename
4097327
Link To Document