Title :
Experience with Ac Calibration of Conventional CCC
Author_Institution :
Electricity Group, MIKES, Espoo
Abstract :
Overview of the sources of dc and ac ratio errors in conventional cryogenic current comparator (CCC) is presented. Preliminary results of the measurements of ac ratio error in overlapped-tube type CCC are given. The measured maximum deviation from the nominal ratio in CCC used in MIKES is less than 2 ppm in frequency range 0-200 Hz
Keywords :
calibration; current comparators; measurement errors; 0 to 200 kHz; AC calibration; MIKES; ac ratio error measurements; cryogenic current comparator; dc ratio errors; Attenuation; Bridges; Calibration; Electrical resistance measurement; Frequency measurement; Geometry; Leakage current; Magnetic materials; Toroidal magnetic fields; Wires;