Abstract
Commercial quantum Hall resistance standard has been put into operation at the CMI. Small, but systematic discrepancy between results obtained at 2nd and 4th QHE plateaux was observed and its possible origin will be discussed. Unpredictable drift of the built-in 100Omega transfer standard will also be dealt with
Keywords
electric resistance measurement; quantum Hall effect; transfer standards; 100 ohm; Czech Metrological Institute; QHE plateaux; commercial quantum Hall resistance standard; transfer standard; Bridge circuits; Calibration; Cryogenics; Helium; Laboratories; Manufacturing; Physics; Resistors; Superconducting magnets; System software;