DocumentCode
3449454
Title
A multi-channel model for myoelectric control
Author
Zhang, Y.T. ; Parker, Philip A. ; Scott, R.N.
Author_Institution
New Brunswick Univ., Fredericton, NB, Canada
fYear
1988
fDate
4-7 Nov. 1988
Firstpage
1133
Abstract
A multimyoelectric channel model, based on muscle physiology, is proposed for myoelectric control. The model performance analysis and simulation show that the multimyoelectric channel has a higher signal-to-noise ratio (SNR) than the conventional myoelectric channel. For the case of the same unit number and the same mean firing rates in groups, the multichannel SNR improvement factor over the single conventional channel is equal to the number of uncorrelated channels.<>
Keywords
biocontrol; bioelectric potentials; muscle; neurophysiology; physiological models; prosthetics; model performance analysis; multi-channel model; multimyoelectric channel; muscle physiology; myoelectric control; powered prostheses; signal-to-noise ratio; simulation; uncorrelated channels;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society, 1988. Proceedings of the Annual International Conference of the IEEE
Conference_Location
New Orleans, LA, USA
Print_ISBN
0-7803-0785-2
Type
conf
DOI
10.1109/IEMBS.1988.94780
Filename
94780