Title :
1000 V DC Voltage Ratio Calibration at VNIIM
Author :
Katkov, Alexander
Author_Institution :
D.I. Mendeleyev Inst. for Metrol., St. Petersburg
Abstract :
To calibrate dc voltage calibrators and voltmeters above 10 V with ratio uncertainty less than 10-6 resistive dividers are used. Method for the calibration of resistive divider at applied voltage up to 1000 V is described. Estimated relative combined uncertainty of 1000 V/10 V was 0.1 ppm (k=2)
Keywords :
calibration; measurement uncertainty; voltage measurement; voltmeters; 1000 V; dc voltage ratio calibration; measurement uncertainty; resistive dividers; voltage measurement; voltmeters; Bridge circuits; Calibration; Circuit testing; Electrical resistance measurement; Joining processes; Metrology; Resistors; Uncertainty; Voltage; Voltmeters;