• DocumentCode
    3449699
  • Title

    Optimizing the Series Connection of Thermal Converters for AC-DC Current Calibration - Experimental Verification

  • Author

    Lind, K. ; Espedalen, J.H. ; Slinde, H.

  • Author_Institution
    Justervesenet
  • fYear
    2004
  • fDate
    38139
  • Firstpage
    583
  • Lastpage
    584
  • Abstract
    Connecting thermal current converters (TCC) in series for calibration of current ac-dc difference, attention has to be paid to the various current leakages that may occur. Calculations indicate leakage error less than 2*10-6 at frequencies up to 100 kHz. Measured values show that error contributions are less than 1*10-6 at 100 kHz, which is well below the predicted value
  • Keywords
    AC-DC power convertors; calibration; electric current measurement; measurement standards; 100 kHz; ac-dc current calibration; leakage current; thermal current converters; Analog-digital conversion; Calibration; Circuits; Computer errors; Frequency estimation; Impedance; Joining processes; Leakage current; Position measurement; Resistance heating;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements Digest, 2004 Conference on
  • Conference_Location
    London
  • Print_ISBN
    0-7803-8494-6
  • Electronic_ISBN
    0-7803-8494-6
  • Type

    conf

  • DOI
    10.1109/CPEM.2004.305390
  • Filename
    4097384