Abstract
When connecting thermal current converters (TCC) in series for calibration of current ac/dc difference, attention has to be paid to the various current leakages that may occur. The series connection has been modelled and a specially designed "T-piece" constructed. Calculations indicate leakage error less than 210-6 in delta at frequencies up to 100 kHz
Keywords
AC-DC power convertors; calibration; electric current measurement; T-piece construction; ac-dc current difference calibration; leakage current; thermal current converters; Analog-digital conversion; Cable shielding; Calibration; Circuits; Frequency; Impedance; Joining processes; Leakage current; Uncertainty; Voltage;