Title :
Fast, Continuous and Accurate Frequency Shift Measurement in the AC Electrogravimetry Technique
Author :
Torres, Róbinson ; Arnau, Antonio ; Perrot, Hubert
Author_Institution :
Antioquia Sch. of Eng. Sci. Inst., Envigado
fDate :
May 29 2007-June 1 2007
Abstract :
In this paper, an electronic system designed to measure the frequency shifts in an experimental AC electrogravimetry setup has been presented. This system is committed to acquire appropriated data for determining the electrogravimetric transfer function (EGTF) and provide information regarding the mass transfer in an electrochemical cell in the AC electrogravimetry technique, but maintaining a good trade-off between the locking frequency bandwidth and the resolution in the frequency tracking. The system is based on an analogue-digital phase locked loop (A-D PLL). The achieved results prove the system reliability to conduct the fast, continuous and accurate frequency shifts measurement under experimental condition in AC electrogravimetry.
Keywords :
crystal resonators; electrochemical analysis; frequency measurement; mass transfer; microbalances; phase locked loops; transfer functions; AC electrogravimetry technique; analogue-digital phase locked loop; electrochemical cell; electrogravimetric transfer function; electronic system; frequency shift measurement; frequency shifts; frequency tracking resolution; locking frequency bandwidth; mass transfer; Bandwidth; Biomedical engineering; Biomedical measurements; Frequency conversion; Frequency measurement; Low pass filters; Phase locked loops; Resonant frequency; Transfer functions; Voltage;
Conference_Titel :
Frequency Control Symposium, 2007 Joint with the 21st European Frequency and Time Forum. IEEE International
Conference_Location :
Geneva
Print_ISBN :
978-1-4244-0646-3
Electronic_ISBN :
1075-6787
DOI :
10.1109/FREQ.2007.4319019