DocumentCode :
3452937
Title :
Frequency pulling in cesium beam frequency standards due to delta M=±1 (sigma) transitions
Author :
Cutler, L.S. ; Flory, C.A. ; Giffard, R.P. ; De Marchi, A.
Author_Institution :
Hewlett-Packard Lab., Palo Alto, CA, USA
fYear :
1991
fDate :
29-31 May 1991
Firstpage :
544
Lastpage :
553
Abstract :
Results of a theoretical investigation of the frequency pulling by sigma transitions and some comparisons with an actual cesium dual-beam tube are presented. It is shown that the sigma transitions adjacent to the clock transition always involve one of the clock levels so their frequency pulling (Ramsey pulling) differs fundamentally from the pulling by pi transitions (Rabi pulling). Rabi pulling is caused by the residual slope due to unbalanced pi transitions and can be significantly reduced by background slope detection techniques. Ramsey pulling results from a fundamental shift in the frequency of the clock transition rather than a background slope. Ramsey pulling can only be reduced by decreasing the sigma transitions and the population asymmetries. A single-beam tube has been designed that has greatly reduced sigma transitions compared with the dual-beam tubes
Keywords :
atomic clocks; caesium; frequency measurement; frequency stability; measurement standards; Cs beam frequency standards; Cs dual beam tube; Rabi pulling; Ramsey pulling; background slope detection; frequency pulling; pi transitions; sigma transitions; sigma transitions reduction; single-beam tube; Atomic beams; Atomic measurements; Clocks; Frequency measurement; Laboratories; Magnetic fields; Microwave magnetics; Optical modulation; Optical pumping; Shape;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Frequency Control, 1991., Proceedings of the 45th Annual Symposium on
Conference_Location :
Los Angeles, CA
Print_ISBN :
0-87942-658-6
Type :
conf
DOI :
10.1109/FREQ.1991.145949
Filename :
145949
Link To Document :
بازگشت