Title :
Full integration of a liver assay panel on a centrifugal microfluidic platform
Author :
Nwankire, Charles E. ; Burger, Robert ; O´Connell, Triona ; Glennon, Thomas ; Ducree, J.
Author_Institution :
Nat. Center for Sensor Res., Dublin City Univ., Dublin, Ireland
Abstract :
In this report we present the full integration and automation including liquid reagent storage of a standard liver assay panel (LAP) on a centrifugal microfluidic platform. This “lab-on-a-disc” platform concurrently runs a full LAP consisting of six enzymatic assays (albumin (ALB), alkaline phosphatase (ALP), gamma glutamyl transferase (GGT), total serum protein (PROT), total (TBIL) and direct (DBIL) bilirubin) from a ~50-μl sample of whole blood. Our technology builds upon our previously reported, merely rotationally actuated water dissolvable film (DF) based valves [1]. In this paper we demonstrate that the sacrificial water DF valves can be independently actuated over an unprecedentedly wide range of spin rates spanning between 10 Hz and 60 Hz! Furthermore, we for the first time demonstrate the sequential actuation of 16 (!) independent DF valves, thus enabling a significantly enhanced level of flow control for the delivery of all LAP reagents. The results, which have been benchmarked against the common assay in a hospital lab, display a low CV of 6.8%, only.
Keywords :
biomedical equipment; blood; centrifuges; enzymes; liver; microfluidics; patient diagnosis; proteomics; ALB; ALP; DBIL; GGT; PROT; TBIL; albumin; alkaline phosphatase; centrifugal microfluidic platform; direct bilirubin; enzymatic assays; gamma glutamyl transferase; lab-on-a-disc platform; liquid reagent storage; liver assay panel; sacrificial water DF valve; total serum protein; whole blood sample; Blood; Films; Hospitals; Liver; Microfluidics; Spinning; Valves; Centrifugal; Dissolvable films; Liver assay; Valves; microfluidics;
Conference_Titel :
Solid-State Sensors, Actuators and Microsystems (TRANSDUCERS & EUROSENSORS XXVII), 2013 Transducers & Eurosensors XXVII: The 17th International Conference on
Conference_Location :
Barcelona
DOI :
10.1109/Transducers.2013.6626781