Title :
Towards a wireless and fully-implantable ECoG system
Author :
Tolstosheeva, E. ; Hoeffmann, J. ; Pistor, Jonas ; Rotermund, D. ; Schellenberg, T. ; Boll, D. ; Hertzberg, T. ; Gordillo-Gonzalez, V. ; Mandon, S. ; Peters-Drolshagen, D. ; Schneider, Markus ; Pawelzik, K. ; Kreiter, A. ; Paul, Sudipta ; Lang, Walter
Author_Institution :
Inst. for Microsensors, -actuators &-Syst., Univ. of Bremen, Bremen, Germany
Abstract :
This paper presents a wireless system, designed for electrocortical (ECoG) neural recordings, consisting of an implantable flex-rigid ECoG array and a wireless electronic platform. The array is designed for implantation on top of the visual cortex of a Rhesus monkey. The electronic platform contains pre-amplifiers and multiplexer stages, analog-digital converters (ADCs), an inhouse developed and reconfigurable ASIC, an RF data transceiver and an inductive energy link. Our ASIC provides user-defined selection of channels (≤128), adjustable channel resolution (1-16bit), sample rate (39S/s-10kS/s), and configurable analog filters. The functionality of the whole system is successfully demonstrated by an in vitro test.
Keywords :
analogue-digital conversion; application specific integrated circuits; bioelectric phenomena; biomedical electronics; biomedical telemetry; neurophysiology; preamplifiers; radio transceivers; ASIC; RF data transceiver; Rhesus monkey; analog-digital converters; electrocortical neural recordings; implantable flex-rigid ECoG array; inductive energy link; multiplexer stages; preamplifiers; visual cortex; wireless electronic platform; wireless fully-implantable ECOG system; Application specific integrated circuits; Arrays; Connectors; Electrodes; Impedance; Polyimides; Wireless communication; Wireless ECoG system; configurable digital ASIC; flex-rigid neural array; local field potentials; wireless RF datalink;
Conference_Titel :
Solid-State Sensors, Actuators and Microsystems (TRANSDUCERS & EUROSENSORS XXVII), 2013 Transducers & Eurosensors XXVII: The 17th International Conference on
Conference_Location :
Barcelona
DOI :
10.1109/Transducers.2013.6626783