DocumentCode :
3455613
Title :
Considerations on the Measurement of the Stability of Oscillators with Frequency Counters
Author :
Dawkins, Samuel T. ; McFerran, John J. ; Luiten, André N.
Author_Institution :
Univ. of Western Australia, Crawley
fYear :
2007
fDate :
May 29 2007-June 1 2007
Firstpage :
759
Lastpage :
764
Abstract :
The most common time-domain measure of frequency stability, the Allan variance, is typically estimated using a frequency counter. Close examination of the operation of modern high-resolution frequency counters shows that they do not make measurements in the way commonly assumed. The consequence is that the results typically reported by many laboratories using these counters are not, in fact, the Allan variance, but a distorted representation. We elucidate the action of these counters by consideration of their operation in the Fourier domain, and demonstrate that the difference between the actual Allan variance and that delivered by these counters can be very significant for some types of oscillators. We also discuss ways to avoid, or account for, a distorted estimation of Allan variance.
Keywords :
Fourier analysis; oscillators; time-domain analysis; Allan variance; Fourier domain; frequency counter; frequency counters; oscillators stability measurement; time-domain measure; Australia; Counting circuits; Distortion measurement; Fluctuations; Frequency estimation; Frequency measurement; Oscillators; Physics; Stability; Time measurement;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Frequency Control Symposium, 2007 Joint with the 21st European Frequency and Time Forum. IEEE International
Conference_Location :
Geneva
ISSN :
1075-6787
Print_ISBN :
978-1-4244-0646-3
Electronic_ISBN :
1075-6787
Type :
conf
DOI :
10.1109/FREQ.2007.4319178
Filename :
4319178
Link To Document :
بازگشت