• DocumentCode
    3459520
  • Title

    Interpolation method for time delay estimation in the RF-signal crosscorrelation technique for blood velocity measurement

  • Author

    Lai, Xinmin ; Torp, H.

  • Author_Institution
    Dept. of Cybern. Eng., Norwegian Univ. of Sci. & Technol., Norway
  • Volume
    2
  • fYear
    1996
  • fDate
    3-6 Nov 1996
  • Firstpage
    1211
  • Abstract
    The cross correlation method (CCM) for blood flow velocity is based on the time delay estimation of the echoes from pulse-to-pulse. The sampling frequency is usually kept low in order to reduce computation complexity, and the peak in the correlation function is found by interpolating the correlation function. The parabolic-fit interpolation method introduces bias at low ratio of sampling rate to ultrasound center frequency. In this work, 4 different interpolation methods are suggested to improve the estimation accuracy. 1. Parabolic-fit with bias-compensation, derived from a theoretical signal model. 2. Parabolic-fit combined with linear filter interpolation to the correlation function. 3. Parabolic-fit to the correlation function of the complex signal envelope. 4. Matched filter applied to the correlation function interpolation. The new interpolation methods are analyzed both by computer simulated signals, and RF-signals recorded from patient data with velocities up to 4 times the Nyquist velocity. The results show that these methods have similar estimation accuracy when signal to noise ratio(SNR) is high and the matched filter interpolation gives the better performance when SNR is low
  • Keywords
    acoustic signal processing; biomedical ultrasonics; blood flow measurement; interpolation; medical signal processing; ultrasonic velocity measurement; Nyquist velocity; RF-signal crosscorrelation technique; blood velocity measurement; complex signal envelope; computation complexity reduction; estimation accuracy improvement; matched filter; parabolic-fit interpolation method; sampling frequency; time delay estimation; ultrasound center frequency; Blood flow; Correlation; Delay effects; Delay estimation; Frequency; Interpolation; Matched filters; Sampling methods; Signal to noise ratio; Ultrasonic imaging;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Ultrasonics Symposium, 1996. Proceedings., 1996 IEEE
  • Conference_Location
    San Antonio, TX
  • ISSN
    1051-0117
  • Print_ISBN
    0-7803-3615-1
  • Type

    conf

  • DOI
    10.1109/ULTSYM.1996.584208
  • Filename