    Analysis of factors influencing the accuracy of 3D Doppler technique: simulations and experimental results

  • Author

    Calzolai, M. ; Capineri, L. ; Fort, A. ; Masotti, L. ; Pasqualis, A. ; Rocchi, S. ; Scabia, M.

  • Author_Institution
    Dept. of Electron. Eng., Florence Univ., Italy
  • Volume
    2
  • fYear
    1996
  • fDate
    3-6 Nov 1996
  • Firstpage
    1245
  • Abstract
    In this work the performance of a prototype three-dimensional (3D) PW Doppler system producing 3D velocity field maps is analyzed by means of measurements on phantoms with known velocity profiles and of simulations. The developed 3-D Doppler system gives a measurement of the velocity vector in a sample volume by means of a probe consisting of four confocal annular array transducers oriented along different directions and three programmable analogue Doppler channels
  • Keywords
    Doppler measurement; biomedical ultrasonics; measurement errors; ultrasonic transducer arrays; ultrasonic velocity measurement; vectors; 3D Doppler technique accuracy; confocal annular array transducers; known velocity profiles; phantom measurements; programmable analogue Doppler channels; sample volume; velocity vector; Analytical models; Design engineering; Imaging phantoms; Performance analysis; Probes; Transducers; Ultrasonic imaging; Ultrasonic variables measurement; Velocity measurement; Volume measurement;