DocumentCode :
3461204
Title :
A modified time-of-flight tomography concept for ultrasonic breast imaging
Author :
Krueger, M. ; Pesavento, A. ; Ermert, H.
Author_Institution :
Dept. of Electr. Eng., Ruhr-Univ., Bochum, Germany
Volume :
2
fYear :
1996
fDate :
3-6 Nov 1996
Firstpage :
1381
Abstract :
Based on a special approach for ultrasonic breast imaging (CARI: clinical amplitude/velocity reconstructive imaging) a modified time-of-flight tomographic imaging algorithm is introduced. Problems of limited angle and signal-to-noise ratio are discussed. Despite fundamental limits phantom experiments show promising results. Compared to CARI the authors´ algorithm leads to an improved lateral resolution and also an axial resolution was obtained, which is missing in CARI images
Keywords :
acoustic tomography; biomedical ultrasonics; image reconstruction; image resolution; medical image processing; axial resolution; clinical amplitude/velocity reconstructive imaging; lateral resolution; limited angle; medical diagnostic imaging; modified time-of-flight tomography concept; signal-to-noise ratio; ultrasonic breast imaging; Acoustic measurements; Apertures; Breast; Brightness; Image resolution; Imaging phantoms; Signal resolution; Signal to noise ratio; Tomography; Ultrasonic imaging;