DocumentCode :
3461472
Title :
Methods for improving image quality in whole body PET scanning
Author :
Dahlbom, Magnus ; Yu, Dan-Chu ; Cherry, Simon R. ; Chatziioannou, Arion ; Hoffman, Edward J.
Author_Institution :
Dept. of Radiol. Sci., California Univ. Sch. of Med., Los Angeles, CA, USA
fYear :
1991
fDate :
2-9 Nov. 1991
Firstpage :
1587
Abstract :
Whole body PET (positron emission tomography) imaging is performed by acquiring data at multiple axial positions. From this data set, coronal and sagittal cross sectional images are formed by reorienting the transaxial tomographic images. Due to the short acquisition time at each axial position the noise levels in the final images are relatively high. The aim of the present work is to optimize some of the scanning parameters for whole body PET imaging to achieve the best possible image quality. It is noted that the detection efficiency of the PET system can be improved by using more coincidence plane combinations in addition to the conventional direct and cross planes. The effect of acquiring an additional set of coincidence planes with a plane separation of +or-2 was studied and showed a reduction in noise and improved uniformity without a significant loss in resolution. The effect of different sampling schemes was also studied. Using a continuous sampling scheme by moving the bed in sets of 3.38 mm results in better image uniformity together with a reduction in noise in comparison with images acquired using the standard interleaved sampling mode.<>
Keywords :
computerised tomography; radioisotope scanning and imaging; acquisition time; continuous sampling scheme; coronal cross sectional images; cross planes; direct planes; image quality improvement methods; medical diagnostic imaging; nuclear medicine; sagittal cross sectional images; scanning parameters optimization; standard interleaved sampling mode; whole body PET scanning; Biomedical imaging; Contracts; Detectors; Image quality; Image sampling; Noise level; Noise reduction; Nuclear medicine; Tomography; Whole-body PET;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium and Medical Imaging Conference, 1991., Conference Record of the 1991 IEEE
Conference_Location :
Santa Fe, NM, USA
Print_ISBN :
0-7803-0513-2
Type :
conf
DOI :
10.1109/NSSMIC.1991.259180
Filename :
259180
Link To Document :
بازگشت