DocumentCode :
3461629
Title :
Comparison of a lock-in amplifier and a digital sampling voltmeter as null detectors for differential measurements between sinusoidal and stepwise approximated ac signals
Author :
Georgakopoulos, D. ; Budovsky, I. ; Palafox, L.
Author_Institution :
Nat. Meas. Inst., Lindfield, NSW, Australia
fYear :
2010
fDate :
13-18 June 2010
Firstpage :
60
Lastpage :
61
Abstract :
This paper presents the results of using two different types of a null detector, a lock-in amplifier and a digital sampling voltmeter, to measure the difference between a sinusoidal and a stepwise approximated AC voltage generated by means of a programmable Josephson voltage standard. Preliminary results of ac-dc difference measurement made on a thermal converter using the two null detectors agree within 0.5 μV/V at 10 Hz and 20 Hz.
Keywords :
amplifiers; convertors; digital voltmeters; measurement systems; ac-dc difference measurement; differential measurements; digital sampling voltmeter; frequency 10 Hz; frequency 20 Hz; lock-in amplifier; null detectors; programmable Josephson voltage standard; sinusoidal approximated ac signals; stepwise approximated ac signals; Analog-digital conversion; Australia; Detectors; Differential amplifiers; Electromagnetic measurements; Instruments; Sampling methods; Synchronization; Voltage; Voltmeters;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Precision Electromagnetic Measurements (CPEM), 2010 Conference on
Conference_Location :
Daejeon
Print_ISBN :
978-1-4244-6795-2
Type :
conf
DOI :
10.1109/CPEM.2010.5543392
Filename :
5543392
Link To Document :
بازگشت