  • Abstract
    Specifications were developed for a PET (positron emission tomography) system that emphasizes high-count-rate performance and is specifically designed for the estimation of myocardial perfusion using oxygen-15 water and dynamic imaging protocols. The seven-slice CsF detector array is a one-on-one configuration, and has a reconstructed FOV (field of view) of 52 cm and an axial FOV of 10.7 cm. Wobble and axial translation are used. The gantry and associated electronics were interfaced to a new computer system, data collection and image reconstruction software was developed and the performance of the system was characterized with phantom, animal, and human studies. The system sensitivity is 136000 cts/s/ mu Ci/cm/sup 3/ and the reconstructed resolution is better than 9 mm. Imaging a 20-cm uniform phantom with 4.3 mu Ci/cm/sup 3/, the system collects a total rate of 1.57 Mcts/s with a loss fraction of 0.72 and a ratio of randoms/trues of 0.64.<>
  • Keywords
    biomedical equipment; cardiology; computerised tomography; radioisotope scanning and imaging; CsF detector array; H/sub 2//sup 15/O; axial translation; computer system; dynamic cardiac imaging; field of view; high-count-rate performance; image reconstruction software; imaging protocols; loss fraction; medical diagnostic imaging; medical instrumentation; myocardial perfusion; nuclear medicine; performance characterization; positron emission tomography; whole body PET system; wobble; Computer interfaces; Detectors; Image reconstruction; Imaging phantoms; Myocardium; Positron emission tomography; Protocols; Sensor arrays; Software performance; Whole-body PET;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium and Medical Imaging Conference, 1991., Conference Record of the 1991 IEEE
  • Conference_Location
    Santa Fe, NM, USA
  • Print_ISBN
    0-7803-0513-2
  • Type

    conf

  • DOI
    10.1109/NSSMIC.1991.259191
  • Filename
    259191