Abstract :
A cylindrically shaped PET (positron emission tomography) camera designed to utilize cross-plane events is being developed. A two-dimensional BGO scintillation detector was designed for the camera. It uses a rectangular array of detector elements and a hex array of PM tubes that are supported by a thin cylindrical light guide-eight detector elements/PM tube. The detector is 60 cm in diameter by 12 cm axially. The BGO elements are 3-mm*5.7-mm*30-mm-deep and form 20 rings of 640 elements. A narrow light distribution is achieved. To decrease the system deadtime, scintillation events are processed using only near neighbor PM tube signals. The goal is to achieve high sensitivity and high resolution in energy, time, and space. The preliminary data obtained from the camera indicate that the design goals will be achieved. Algorithms that equalize the PM tube gains will improve energy resolution, spatial resolution, and stability.<>
Keywords :
biomedical equipment; computerised tomography; radioisotope scanning and imaging; 12 cm; 2D BGO scintillation detector; 3 mm; 30 mm; 5.7 mm; 60 cm; algorithms; cross-plane events; cylindrical PET detector characteristics; detector elements; energy resolution; hex array; light distribution; medical diagnostic imaging; medical instrumentation; nuclear medicine; positron emission tomography; rectangular array; scintillation events; spatial resolution; stability; system deadtime; thin cylindrical light guide; Cameras; Energy resolution; Optical arrays; Positron emission tomography; Scintillation counters; Sensor arrays; Signal processing; Signal resolution; Solid scintillation detectors; Spatial resolution;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium and Medical Imaging Conference, 1991., Conference Record of the 1991 IEEE
Conference_Location :
Santa Fe, NM, USA
Print_ISBN :
0-7803-0513-2
Type :
conf
DOI :
10.1109/NSSMIC.1991.259193
Filename :
259193
Link To Document :
بازگشت