• DocumentCode
    3463223
  • Title

    Capability evaluation of a sinogram error detection and correction method in computed tomography

  • Author

    Huang, Sung-Cheng ; Yu, Dan-Chu

  • Author_Institution
    Sch. of Med., California Univ., Los Angeles, CA, USA
  • fYear
    1991
  • fDate
    2-9 Nov. 1991
  • Firstpage
    2073
  • Abstract
    An iterative method for detecting and correcting errors in sinograms is described and its capability is evaluated. The method is based on the use of a reconstruction-reprojection procedure to check the consistency of a sinogram. Adjustments of the sinogram are made iteratively to satisfy gradually the consistency requirement. The effectiveness of the method for detecting and correcting errors due to scanner detector problems is demonstrated with computer-simulated data and with real data from positron emission tomography. The potential usefulness of the method for quality control of scanners is discussed.<>
  • Keywords
    computerised tomography; image reconstruction; radioisotope scanning and imaging; iterative adjustments; positron emission tomography; reconstruction-reprojection procedure; scanner detector problems; scanner quality control; sinogram correction method; sinogram error detection; Computational modeling; Computed tomography; Computer errors; Computer simulation; Detectors; Error correction; Image reconstruction; Nuclear medicine; Positron emission tomography; Quality control;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium and Medical Imaging Conference, 1991., Conference Record of the 1991 IEEE
  • Conference_Location
    Santa Fe, NM, USA
  • Print_ISBN
    0-7803-0513-2
  • Type

    conf

  • DOI
    10.1109/NSSMIC.1991.259280
  • Filename
    259280